Classification
Protection Class: Type BF defibrillation proof
Safety: IEC 60601-1
Patient Range
Adult and Pediatrics or neonatal (optional)
NIBP Measurement
Method: Oscillometric
Operation Modes: Manual/Automatic/STAT
Measurement Unit: mmHg/kPa selectable
Measurement Typeset: Systolic, Diastolic, and Mean
Measurement Range Infant Adult
SYS 30-135 30-255
DIA 15-110 15-220
MAP 20-125 20-345
Over-pressure protection: YES
Resolution: 1 mmHg
SpO2
Display range: 0 - 100%
Accuracy: 80% - 100%:
62%
70% - 79%:
63%
0 - 69%: Unspecified
Pulse Rate
Measurement Range: 30 - 235bpm
Resolution: 1bpm
Accuracy: 30 - 100:
62bpm
101 - 235:
62%
0 - 29: Unspecified
Display
Type: LCD & LED
Alarm
Alarm: NIBP (SYS, DIA, MAP), SpO2, PR, Probe off
Vital Signs Monitor
Finger out, Low voltage
Modes: 3-level Audio & Visual alarm, Error code
Appearance
Dimension: 296mmX166mmX96mm
Weight: 1.8kg (including battery)
Environmental
Power: AC100 - 230V, 50/60Hz (
63Hz)
Operating temperature: 5°C - 40°C
Storage temperature: -20°C - 55°C
Operation Humidity:
≤80%, non-condensing
Storage Humity: 0 - 93%, non-condensing
Storage
Patient ID: 1 - 99
Data records: up to 72 hours
Data transmission (Optional)
Data cable through USB interface
Flash Disk
Wired LAN to choice central monitoring system
Battery
Type: Built-in rechargeable battery
Operation time: more than 2 hours (continuous
operation)
Printer (Optional)
Type: built-in thermal recorder
Width of printing paper: standard 50(+0/-1)mm |